Hygoclave 90

Dimensions (H x B x T): 47,5 cm x 54 cm x 52 cm
Weight: 65 kg
Chamber volume: 19 l
Internal dimensions, useable volume (∅ x T): 26 cm x 35 cm
Ambient operational temperature: 15 – 35 °C
Maximum power consumption: 2600 W
Maximum number of trays: 6
Maximum load solid: 6 kg (10 kg for containers)
Maximum load porous: 2 kg
Maximum load container 65 mm: 3
Noise level*: <65 dB(A)
Fresh water and waste water tank volume: 4,5 l
Meets the guidelines and standards: 93/42/EWG (Klasse IIb), EN 13060:2015, EN 61010-1, EN 61010-2-040, EN 61326-1

Toote põhiinfo

Program Operating time Drying Load Test Type
Universal 134 °C
5,5 min/12 – 23 min*
16 min For 6 kg of instruments or 2 kg of textiles, in single and multiple packaging. For 10 kg with container loading.** Helix-Test B
Quick 134 °C
5,5 min/10 – 17 min*
10,5 min For instruments up to 1.5 kg single packaging or 6 kg unpackaged. Helix-Test B
Gentle 121 °C
20,5 min/26 – 41 min*
16 min For 6 kg of instruments or 2 kg of textiles, in single and multiple packaging, especially for sterilisation items. Helix-Test B
Prion 134 °C
20 min/26 – 37 min*
16 min For 6 kg of instruments or 2 kg of textiles, in single and multiple packaging, especially in cases of suspected infection with prions. For 10 kg with container loading.** Helix-Test B